 Agilent N4376B 20 GHz and 26.5 GHz Multimode Lightwave Component Analyzer
Data Sheet
www..com
General Information
Agilent's N4376B Lightwave Component Analyzer (LCA) is the instrument of choice to test short wavelength 10G Ethernet, Fibre Channel FCx8, FCx10 and FCx16 electro-optical components, with up to 20 or 26.5 GHz modulation range. The N4376B also supports the test of transmitter and receivers for optical computer backplanes and optical chip-to-chip connections in high speed computers and server applications. Modern optical transmission and datacom systems require fast, accurate and repeatable characterization of the core electro-optical components. These core subcomponents (lasers, modulators and detectors) have significant impact on the performance of the transmitter and the receiver with respect to modulation bandwidth, jitter, gain, and distortion of the final transceiver. For frequency dependent responsivity measurements the N4376B extends opto-electronic S-parameter measurements to multimode devices in the 850 nm wavelength range, which was not possible with the standard 8703A/B LCA 8703A/B. With a completely new design of the optical test set and a new RF-switched architecture, together with the latest PNA family of network analyzers, the N4376B guarantees excellent electro-optical measurement performance. In addition a unique new calibration concept significantly reduces time from powering up the LCA until the first calibrated measurement can be made. This increases productivity in R&D and on the manufacturing floor. The fully integrated "turnkey" solution reduces time to market, compared to the time-consuming development of a selfmade setup. By optimizing the electrical and optical design of the N4376B for lowest noise and ripple, the accuracy has been improved by more than a factor of 3 and is now independent of the electrical reflection coefficient of the device under test. It's the excellent accuracy that improves the yield from tests performed with the N4376B, by narrowing margins needed to pass the tested devices. Traceability ensures worldwide comparability of test results. The advanced optical design together with temperaturestabilized transmitter and receiver ensures repeatable measurements over days without recalibration. Using the advanced measurement capabilities of the network analyzer, all S-parameter related characteristics of the device under test, like responsivity, ripple and 3dB-cutoff frequency, can be qualified with the new N4376B Lightwave Component Analyzer from 10 MHz to 20/26.5 GHz.
www..com
Key benefits
Traceable multimode S21 test, right at 850 nm target wavelength. IEEE 802.3ae launched power distribution leads to test results comparable to the final application Fast and easy measurement setup and calibration for all standard tests High confidence and fast time-to-market with a traceable turnkey solution. Significantly increased productivity using the fast and easy measurement setup with an unique new calibration process leads to lower cost of test. Test right at target launch condition eliminates test uncertainty Identical LCA software and remote control across the N437xB family simplifies integration Relative frequency response uncertainty @ 20GHz: 1.5 dB 1.0 dB (typical) Absolute frequency response uncertainty @ 20GHz: 2.0 dB (typ.)for E/O measurements 1.8 dB (typ.)for O/E measurements Noise floor @ 20GHz: -69 dB W/A for E/O measurements -68 dB A/W for O/E measurements Transmitter wavelength: 850 nm 10 nm Supported connectors LC or SC straight Built-in optical power meter For fast transmitter power verification Powerful remote control: State of the art programming interface based on Microsoft .NET or COM. Warranty: 1 year warranty is standard for N4376B Lightwave Component Analyzer; Extension to 3 years available.
The network analyzer
The N4376B comes in two basic versions. The economic line is based on a PNA-L network-analyzer and is available as a 2 port system. The high end version is based on the new PNAX and, extended optical modulation index (OMI) and is available with 2 or4 ports. The PNA-X based LCA is calibrated up to 26.5 GHz.
2
Agilent N4376B Applications
In digital photonic transmission systems, the performance is ultimately determined by bit error ratio test (BERT), which describes the performance of the whole system. However it is necessary to design and qualify subcomponents like modulators and receivers, which are analog by nature, with different parameters. Those parameters are core to the overall system performance. These electro-optical components significantly influence the overall performance of the transmission system via the following parameters: 3dB bandwidth of the electro-optical transmission Relative frequency response, quantifying the electrooptical shape of the conversion. Absolute frequency response, relating to the conversion efficiency of signals from the input to the output, or indicating the gain of a receiver. Electrical reflection at the RF port Only a careful design of these electro-optical components over a wide modulation signal bandwidth guarantees successful operation in the transmission system.
Agilent N4376B Features
Turnkey solution
In today's highly competitive environment, short time-tomarket with high quality is essential for new products. Instead of developing a home-grown measurement solution which takes a lot of time and is limited in transferability and support, a fully specified and supported solution helps to focus resources on faster development and on optimizing the manufacturing process. In the N4376B all optical and electrical components are carefully selected and matched to each other to minimize noise and ripple in the measurement traces. Together with the temperature stabilized environment of the core components, this improves the repeatability and the accuracy of the overall system. Extended factory calibration data at various optical power levels ensures accurate and reliable measurements that can only be achieved with an integrated solution like the N4376B.
Easy calibration
An LCA essentially measures the conversion relation between optical and electrical signals. This is why user calibration of such systems can evolve into a time consuming task. With the new calibration process implemented in the N4376B, the tasks that have to be done by the user are reduced to one pure electrical calibration. The calibration with an electrical microwave calibration module is automated and needs only minimal manual interaction.
Electro-optical components
The frequency response of amplified or unamplified detector diodes, modulators and directly modulated lasers typically depends on various parameters, like bias voltages, optical input power, operating current and ambient temperature. To determine the optimum operating point of these devices, an LCA helps by making a fast characterization of the electrooptic transfer function while optimizing these operating conditions. In parallel the LCA also measures the electrical return loss. In manufacturing it is important to be able to monitor the processes regularly to keep up the throughput and yield. In this case the LCA is the tool of choice to monitor transmission characteristics and absolute responsivity of the manufactured device. The remote control of the N4376B offers another tool to improve the productivity by making automated measurements and analysis of the measured data.
Built-in performance verification
Sometimes it is necessary to make a quick verification of the validity of the calibration and the performance of the system. The N4376B's unique calibration process allows the user to perform a self-test without external reference devices. This gives full confidence that the system performance is within the user's required uncertainty bands.
State-of-the-art remote control
Testing the frequency response of electro-optical components under a wide range of parameters, which is often necessary in qualification cycles, is very time consuming. To support the user in minimizing the effort for performing this huge number of tests, all functions of the LCA can be controlled remotely via LAN over the state-of-the-art Microsoft .NET or COM interface. Based on programming examples for VBA with Excel, Agilent VEE and C++, it is very easy for every user to build applications for their requirements. These examples cover applications like integration of complete LCA measurement sequences.
Electrical components
Electrical components such as amplifiers, filters and transmission lines are used in modern transmission systems and require characterization to ensure optimal performance. Typical measurements are bandwidth, insertion loss or gain, impedance match and linearity. The new switched architecture offers direct access to the electrical outputs and inputs of the network-analyzers just by selecting electrical- to electriwww..com cal measurement mode in the LCA user interface.
3
Integrated optical power meter
In applications where optical power dependence characterization is needed, the average power meter can be used to set the exact average output power of the LCA transmitter by connecting the LCA optical transmitter output, optionally through an optical attenuator, to the LCA optical receiver input. By adjusting the transmitter output power in the LCA user interface or the optical attenuation, the desired transmitter optical power can be set. In cases where an unexpectedly low responsivity is measured from the device under test, it is very helpful to get a fast indication of the CW optical power that is launched into the LCA receiver. The cause might be a bad connection or a bent fiber in the setup. For this reason too, a measurement of the average optical power at the LCA receiver is very helpful for fast debugging of the test setup.
Selectable output power of the transmitter
Most PIN diodes and receiver optical subassemblies (ROSA's) need to be characterized at various average optical power levels. In this case it is necessary to set the average input power of the device under test to the desired value. The variable average optical output power of the LCA transmitter offers this feature. Together with an external optical attenuator, this range can be extended to all desired optical power levels.
Large signal measurements
LCA S21 measurements are typically small-signal linear transfer function measurements. If an electro-optical component must be tested under large signal conditions, normal balanced measurements might lead to wrong measurement results. The PNA-X based version of the LCA offers true balanced measurements for differential ports by offering two independent high power RF sources. With this setup the LCA measures the correct S21 transfer function of E/O components, even in the nonlinear regime. To stimulate O/E components like PIN-TIA receivers under optical large signal conditions, the PNA-X based LCA offers a variable optical modulation index > 50%.
IEEE 802.3ae multimode launch condition
Multimode measurements are typical much more critical regarding repeatability and stability than single mode measuremts. To minimize these effects it is necessary to have well defined and stable mode filling of the transmitter fiber. The N4376B has typical multimode launch conditions or power-distribution in the transmitter fiber as defined by the www..com IEEE 802.3ae standard. The IEEE 802.3ae power-distribution compliance of the N4376B transmitter leads to application realistic and repeatable test results.
4
Definitions
Generally, all specifications are valid at the stated operating and measurement conditions and settings, with uninterrupted line voltage.
Explanation of terms
Responsivity
For electro-optical devices (e.g. modulators ) this describes the ratio of the optical modulated output signal amplitude compared to the RF input amplitude of the device. For opto-electrical devices (e.g. photodiodes) this describes the ratio of at the RF amplitude at the device output to the amplitude of the modulated optical signal input. .
Specifications (guaranteed)
Describes warranted product performance that is valid under the specified conditions. Specifications include guard bands to account for the expected statistical performance distribution, measurement uncertainties changes in performance due to environmental changes and aging of components.
Relative frequency response uncertainty
Describes the maximum deviation of the shape of a measured trace from the (unknown) real trace. This specification has strong influence on the accuracy of the 3-dB cut-off frequency determined for the device under test.
Typical values (characteristics)
Characteristics describe the product performance that is usually met but not guaranteed. Typical values are based on data from a representative set of instruments.
Absolute frequency response uncertainty
Describes the maximum difference between any amplitude point of the measured trace and the (unknown) real value. This specification is useful to determine the absolute responsivity of the device versus modulation frequency.
General characteristics
Give additional information for using the instrument. These are general descriptive terms that do not imply a level of performance.
Frequency response repeatability
Describes the deviation of repeated measurement without changing any parameter or connection relative to the average of this measurements.
Minimum measurable frequency response
Describes the average measured responsivity when no modulation signal is present at the device under test. This represents the noise floor of the measurement system.
Definition of LCA input and output names
www..com
LCA electrical port A
LCA electrical port B
LCA optical output
LCA optical input
5
Measurement capabilities
3dB cut-off frequency (S21), Responsivity (S21), Electrical reflection (S11 or S22), Insertion Loss (IL), Transmission bandwidth, all electrical S-parameter measurements. Target test devices Transmitter (E/O)
Agilent N4376B Specifications
Measurement conditions Modulation frequency range from 10 MHz to 20.0 GHz Foreward RF power +3 dBm Reverse RF power 0 dBm 100 Hz IFBW ("Reduce IF bandwidth at low frequency" enabled) with modulation frequency step size 10 MHz and measurement points on a 10 MHz raster (if not differently stated) Network analyzer set to "stepped sweep - sweep moves in discrete steps" All network-analyzer ports configured in standard coupler configuration ("CPLR ARM" to "RCVB B in", "SOURCE OUT" to "CPLR THRU") After full two-port electrical calibration using an Electronic Calibration Module, Agilent N4691B, at constant temperature (1 C) Modulator bias optimization set to "continous sweep" Measurement frequency grid equals electrical calibration grid DUT signal delay 0.1/IF-BW Specified temperature range: +20 C to +26 C. After warm-up time of 90 minutes Using high quality electrical and optical connectors and RF cables in perfect condition 50 m FC/APC to FC/PC patchcord at the input and output Launched power distribution according to IEEE 802.3ae - 2002, see fig 1 Test performed using an optical reference source with return loss better 45 dB, spectral width FWHM < 10MHz and InGaAs detector
Receiver (O/E) PIN diodes Avalanche photodiodes (APD) Receiver optical subassemblies (ROSA) Optical (O/O) Passive optical components Optical multimode fibers Optical transmission systems
100 90 80
percent power in radius [%]
70 60 50 40
30 www..com 20 10 0 0 5 10 15 20 25
Fig 1; IEEE 802.3ae launch conditions measured with 3 examples
6
Transmitter and Receiver Specifications
Optical Test set Operation frequency range Connector type (optical testset) optical input optical output RF LCA optical input Operating input wavelength range Maximum linear average input power Maximum safe average input power Optical return loss (typ.) [f1] Average power measurement range[f1] Average power measurement uncertainty (typ.) [f1] LCA optical output Optical modulation index (OMI) at 10 GHz (typ.) Output wavelength Lauched Power Distribution (typical) Average output power range Average output power uncertainty (typ.) Average output power stability, 15 minutes (typ.)
[f1] Wavelength within range as specified for LCA optical output [f2] After modulator optimization
[f2] [f1]
Option -322, -382 10 MHz to 20 GHz
Option -312, -314, -392, -394 10 MHz to 26.5 GHz
62.5 m MMF angled with Agilent versatile connector interface 50 m MMF angled with Agilent versatile connector interface 3.5 mm male
750 nm to 1650 nm Optical input: Optical input: > 14 dB Optical input: 0.7 dBo -25 dBm to -1 dBm -1 dBm +3 dBm
25 % @ +3 dBm RF (typ.) 31 % @ +5 dBm RF (typ.) (850 10) nm according to IEEE 802.3ae - 2002 -5 dBm to -1 dBm 0.7 dBo 0.5 dBo
Specifications for electrical-electrical measurements (E/E mode)
For detailed specification of the network analyzer see corresponding data sheet. N4376B: option -322, -382 N5230C -225 option -312, -392 N5242A -200 option -314, -394 N5242A -400 Optical test set Electrical loss of optical test set < 2.0 dBe (typ.)
www..com
7
Specifications for electro-optical measurements at 850 nm
N4376B system with network analyzer N5230C -225 N5242A -200 N5242A -400
(E/O mode)
Specifications are valid under the stated measurement conditions. For wavelength: (850 10) nm System performance Relative frequency response uncertainty DUT response -26 dB(W/A) [f1] -36 dB(W/A) -46 dB(W/A) Absolute frequency response uncertainty (typ.) Frequency response repeatability (typ.) DUT response 26 dB(W/A) [f1] DUT response -26 dB(W/A) [f1] -36 dB(W/A) Minimum measurable frequency response (noise floor ) [f2] [f4]
[f1] [f2] [f4] For DUT optical peak output power +0 dBm. IFBW = 10 Hz. Note: average value over frequency range.
0.05 GHz to 0.2 GHz 1.0 dBe typ. 1.0 dBe typ. 1.1 dBe typ. 2.1dBe
0.2 GHz to 10 GHz 1.3 dBe (0.9 dBe typ) 0.9 dBe typ. 0.9 dBe typ. 2.0 dBe .
10 GHz to 20 GHz 1.5 dBe (1.0 dBe typ) 1.0 dBe typ. 1.3 dBe typ. 2.0 dBe
0.1 dBe 0.15 dBe -65 dB(W/A)
0.1 dBe 0.1 dBe -82 dB(W/A)
0.1 dBe 0.15 dBe -69 dB(W/A)
www..com
8
Specifications for opto-electrical measurements at 850 nm
N4376B system with network analyzer N5230C -225 N5242A -200 N5242A -400
(O/E mode)
Specifications are valid under the stated measurement conditions. For wavelength: (850 10) nm System performance Relative frequency response uncertainty DUT response -21 dB(A/W) [f1] -31 dB(A/W) -41 dB(A/W) Absolute frequency response uncertaint (typ.) Frequency response repeatability (typ.) DUT response 21 dB(A/W) [f1] DUT response -21 dB(A/W) [f1] -31 dB(A/W) Minimum measurable frequency response (noise floor ) [f3] [f4]
[f1] [f2] [f3] [f4] For DUT response max +15 dB (A/W) Output power set to -1 dBm IFBW = 10 Hz. Note: average value over frequency range
0.05 GHz to 0.2 GHz 1.0 dBe typ. 1.0 dBe typ. 1.2 dBe typ. 1.9 dBe
0.2 GHz to 10 GHz 1.3 dBe (0.9 dBe typ) 0.9 dBe typ. 0.9 dBe typ. 1.7 dBe
10 GHz to 20 GHz 1.5 dBe (1.0 dBe typ) 1.1 dBe typ. 1.5 dBe typ. 1.8 dBe
0.25 dBe 0.3 dBe -58 dB(A/W)
0.1 dBe 0.1 dBe -77 dB(A/W)
0.2 dBe 0.25 dBe -68 dB(A/W)
www..com
9
Specifications for optical-optical measurements at 850 nm
N4376B system with network analyzer N5230C -225 N5242A -200 N5242A -400
(O/O mode)
Specifications are valid under the stated measurement conditions. For wavelength: (850 10) nm System performance Relative frequency response uncertainty[f2] Absolute frequency response uncertainty (typ.) Frequency response repeatability (typ.) DUT response -10 dBe [f1] [f2] ( -5.0 dBo) DUT response -10 dBe [f1][f2] ( -5 dBo) DUT response -10 dBe [f1] [f2] ( -5 dBo) 0.15 dBe -53 dBe (-26.5 dBo) 0.1 dBe -70 dBe (-35 dBo) 0.15 dBe -44 dBe (-22 dBo) 1.1 dBe 1.0 dBe 1.0 dBe 0.5 dBe typ. (0.25 dBo) 0.4 dBe (0.2 dBo) 0.5 dBe (0.25 dBe) 0.05 GHz to 0.2 GHz 0.2 GHz to 10 GHz 10 GHz to 20 GHz
Minimum measurable frequency response (noise floor ) [f2] [f3] [f4]
[f1] [f2] [f3] [f4] For DUT response max. 0 dB Average power from LCA optical output set to -1 dBm. IFBW = 10 Hz. Note: average value over frequency range.
www..com
10
General Characteristics
Assembled dimensions: (H x W x D) Option -312, -314 41.3 cm x 43.8 cm x 53.8 cm, (16.3 in x 17.3 in x 21.2 in) -322 41.3 cm x 43.8 cm x 47.3 cm, (16.3 in x 17.3 in x 18.7 in) Weight Product net weight: -312 36 kg (79.4 lbs) -314 46 kg (101.4 lbs) -322 34 kg (74.9 lbs) Packaged product: -312 56 kg (123.5 lbs) -314 66 kg (145.7 lbs) -322 54 kg (119 lbs) Power Requirements 100 to 240 V~, 50 to 60 Hz 2 power cables N5230C max. 350 VA N5242A max 450 VA Optical test set: max. 40 VA Network-analyzer Option 312 N5242A -200 Option 314 N5242A -400 Option 322 N5230C- 225 Storage temperature range -40 C to +70 C Operating temperature range +5 C to +32 C Humidity 15 % to 80 % relative humidity, non-condensing Altitude (operating) 0 ... 2000 m Recommended re-calibration period 1 year
www..com
Shipping contents 1x Network-analyzer depending on option selected 1x N4376B optical test set 2x 81000 NI optical adaptor 1x 4376B-90A01 Getting started 1x 4375B-90CD1 LCA support CD 1x 1150-7896 Keyboard 1x 1150-7799 Mouse 1x 8121-1242 USB cable 1x E5525-10285 UK6 report 1x N4373-61627 electrical short cut cable 1x 9320-6677 RoHS addendum for Photonic T&M accessories 1x 9320-6654 RoHS addendum for Photonic T&M products Additional, option dependent shipping contents: -023 2x LC 50m to FC/APC 0.5 m patch cord -024 2x LC 62.5m to FC/APC 0.5 m patch cord -025 2x SC 50m to FC/APC 0.5 m patch cord -026 2x SC 62.5m to FC/APC 0.5 m patch cord -312, 322 2 port LCA: 1x E7342-60004 0.5 m (m) to (f) high performance RF cable -314 4 port LCA: 2x E7342-60004 0.5 m (m) to (f) high performance RF cable LCA connector types [1] Optical test set LCA port A LCA port B LCA optical input LCA optical output
[1]
3.5 mm (m) 3.5 mm (m) 62.5 m single-mode angled [1], with Agilent universal adapter 50 m single-mode angled[1], with Agilent universal adapter
The optical test set always has angled connectors. For input and output a 50 m or 62.5 m angled to straight LC or SC patchcord must be selected The connection to the DUT is always either LC or SC straight. The jumper cable must always be used in front of the optical testset to protect the connectors of the optical testset. Laser Safety Information All laser sources listed above are classified as Class 1M according to IEC 60825 1 (2001). All laser sources comply with 21 CFR 1040.10 except for deviations pursuant to Laser Notice No. 50, dated 2001-July-26.
11
Mechanical Outline Drawings, option -322, -382 (all dimensions in mm)
www..com
12
Mechanical Outline Drawings, option -312, -314, -392, -394 (all dimensions in mm)
www..com
13
Ordering informations
The N4376B consists of an optical test set and an electrical network analyzer which are mechanically connected. To protect your network analyzer investment, Agilent offers the integration of an already owned PNA-L or PNA-X with the optical test set as listed below. All systems have 1 year warranty. N4376B LCA ordering options Network-analyzer options N4376B - 312 N4376B - 314 N4376B - 322 Network-analyzer integration options N4376B - 382 N4376B - 392 Integration of customer PNA-L - N5230A/C -220, -225 Integration of customer PNA- X - N5242A -200, - N5242A -219 (all specifications typical, max Bias-T voltage 7V, max current 200mA) Integration of customer PNA- X - N5242A -400, - N5242A -419 (all specifications typical, max Bias-T voltage 7V, max current 200mA) 20 GHz 2 port LCA based on N5242A -200 20 GHz 4 port LCA based on N5242A -400 20 GHz 2 port LCA based on N5230C -225
N4376B - 394
Wavelength options N4376B - 103 Other options N4376B - 010 N4376B - 023 N4376B - 024 N4376B - 025 N4376B - 026 Service and Repair R1280A R1282A Required accessories (to be ordered separately) N4691B 2 port microwave electrical calibration module ( -00F recommended) 1 year Return-to-Agilent warranty extended to 3 years Agilent calibration up front support plan 3 year coverage Time domain LC 50m connector interface (external 0.75 m patch cord) LC 62.5m connector interface (external 0.75 m patch cord) SC 50m connector interface (external 0.75 m patch cord) SC 62.5m connector interface (external 0.75 m patch cord) 850 nm optical testset
www..com
14
This page is intentionally left blank
www..com
15
Optical instruments online information
Optical test instruments
www.agilent.com/find/oct
Remove all doubt
Our repair and calibration services will get your equipment back to you, performing like new, when promised. You will get full value out of your Agilent equipment throughout its lifetime. Your equipment will be serviced by Agilent-trained technicians using the latest factory calibration procedures, automated repair diagnostics and genuine parts. You will always have the utmost confidence in your measurements. Agilent offers a wide range of additional expert test and measurement services for your equipment, including initial start-up assistance onsite education and training, as well as design, system integration, and project management. For more information on repair and calibration services, go to
www.agilent.com
For more information on Agilent Technologies' products, applications or services, please contact your local Agilent office. The complete list is available at:
Lightwave Component Analyzers
www.agilent.com/find/lca
www.agilent.com/find/contactus Americas Canada Latin America United States Asia Pacific Australia China Hong Kong India Japan Korea Malaysia Singapore Taiwan Thailand
Polarization solutions
www.agilent.com/find/pol
Spectral analysis products
www.agilent.com/comms/octspectral
(877) 894-4414 305 269 7500 (800) 829-4444
Electro-optical converters
www.agilent.com/find/ref
Optical test instruments accesories Firmware and driver download
www.agilent.com/comms/octfirmware
www.agilent.com/comms/oct-accessories www.agilent.com/find/removealldoubt
Agilent Email Updates
www.agilent.com/find/emailupdates
Agilent photonic discussion forum
www.agilent.com/find/photonic_forum
Get the latest information on the products and applications you select.
1 800 629 485 800 810 0189 800 938 693 1 800 112 929 0120 (421) 345 080 769 0800 1 800 888 848 1 800 375 8100 0800 047 866 1 800 226 008
Agilent Direct
www.agilent.com/find/agilentdirect
Quickly choose and use your test equipment solutions with confidence.
Agilent Open
Europe & Middle East Austria 0820 87 44 11 Belgium 32 (0) 2 404 93 40 Denmark 45 70 13 15 15 Finland 358 (0) 10 855 2100 France 0825 010 700*
*0.125 /minute
Germany
01805 24 6333**
**0.14 /minute
www.agilent.com/find/open Agilent Open simplifies the process of connecting and programming test systems to help engineers design, validate and manufacture electronic products. Agilent offers open connectivity for a broad range of system-ready instruments, open industry software, PC-standard I/O and global support, which are combined to more easily integrate test system development.
Ireland 1890 924 204 Israel 972-3-9288-504/544 Italy 39 02 92 60 8484 Netherlands 31 (0) 20 547 2111 Spain 34 (91) 631 3300 Sweden 0200-88 22 55 Switzerland 0800 80 53 53 United Kingdom 44 (0) 118 9276201 Other European Countries: www.agilent.com/find/contactus
Revised: March 27, 2008
www..com
Product specifications and descriptions in this document subject to change without notice. (c) Agilent Technologies, Inc. 2008 Printed in USA, Rev 1.0, May 16th, 2008 5989-6590EN
www.lxistandard.org LXI is the LAN-based successor to GPIB, providing faster, more efficient connectivity. Agilent is a founding member of the LXI consortium.